DC/DC Converter
TES 2M Series, 2 Watt
•
Supplementary and reinforced
insulation
•
I/O isolation 4000 VACrms rated for 300
Vrms working voltage
•
•
•
•
•
•
Unregulated device
•
•
Available in tape & reel package
2 x MOOP Medical safety
ES 60601-1 IEC 60601-1
UL 62368-1 IEC 62368-1
Industrial safety to UL/IEC/EN 62368-1
Ultracompact SMD-package
Operating temp. range –25°C to +80°C
Qualified for leadfree reflow solder
process
3-year product warranty
The TES 2M series is range of compact 2W DC/DC-converters providing a high
I/O-isolation voltage of 4000 VAC. With a reinforced I/O-isolation system this
product is an economical solution for many applications in instrumentation, industrial controls, medical equipment and everywhere where supplementary- or
reinforced insulation is required. These converters are qualified for high solder
temperature profiles in leadfree solder processes. For automated SMD production lines the devices can be supplied in tape & reel package. Full SMD-design
with exclusive use of ceramic capacitors ensure a very high reliability and a
long product lifetime.
Models
Order Code
Input Voltage
Range
TES 2-0511M
TES 2-0512M
TES 2-0513M
TES 2-0522M
TES 2-0523M
TES 2-1211M
TES 2-1212M
TES 2-1213M
TES 2-1222M
TES 2-1223M
TES 2-2411M
TES 2-2412M
TES 2-2413M
TES 2-2422M
TES 2-2423M
www.tracopower.com
4.5 - 5.5 VDC
(5 VDC nom.)
10.8 - 13.2 VDC
(12 VDC nom.)
21.6 - 26.4 VDC
(24 VDC nom.)
Output 1
Output 2
Vnom
Imax
5 VDC
12 VDC
15 VDC
+12 VDC
+15 VDC
5 VDC
12 VDC
15 VDC
+12 VDC
+15 VDC
5 VDC
12 VDC
15 VDC
+12 VDC
+15 VDC
400 mA
165 mA
133 mA
83 mA
66 mA
400 mA
165 mA
133 mA
83 mA
66 mA
400 mA
165 mA
133 mA
83 mA
66 mA
September 20, 2023
Vnom
Efficiency
Imax
–12 VDC
–15 VDC
83 mA
66 mA
–12 VDC
–15 VDC
83 mA
66 mA
–12 VDC
–15 VDC
83 mA
66 mA
typ.
66 %
66 %
66 %
72 %
73 %
66 %
66 %
66 %
74 %
75 %
66 %
66 %
66 %
74 %
75 %
Page 1 / 4
TES 2M Series, 2 Watt
Input Specifications
Input Current
- At no load
5 Vin models:
12 Vin models:
24 Vin models:
5 Vin models:
12 Vin models:
24 Vin models:
5 Vin models:
12 Vin models:
24 Vin models:
5 Vin models:
12 Vin models:
24 Vin models:
- At full load
Surge Voltage
Recommended Input Fuse
Input Filter
90 mA typ.
40 mA typ.
30 mA typ.
580 mA typ.
240 mA typ.
120 mA typ.
9 VDC max. (1 s max.)
18 VDC max. (1 s max.)
30 VDC max. (1 s max.)
1'000 mA (slow blow)
500 mA (slow blow)
200 mA (slow blow)
(The need of an external fuse has to be assessed
in the final application.)
Internal Capacitor
Output Specifications
Voltage Set Accuracy
Regulation
- Input Variation (1% Vin step)
Ripple and Noise
Capacitive Load
- Load Variation
- Voltage Balance
(symmetrical load)
- 20 MHz Bandwidth
- single output
single output models:
dual output models:
See application note:
dual output models:
5 Vout models:
12 Vout models:
15 Vout models:
12 / -12 Vout models:
15 / -15 Vout models:
- dual output
Minimum Load
Temperature Coefficient
Start-up Time
Short Circuit Protection
±4% max.
1.5% max.
1.5% max.
www.tracopower.com/overview/tes2m
1% max.
150 mVp-p max.
330 µF max.
330 µF max.
330 µF max.
100 / 100 µF max.
100 / 100 µF max.
2 % of Iout max.
(Operation at lower load will not damage the
converter, but it may not meet all specifications)
±0.02 %/K max.
500 ms max.
Limited 0.5 s max., Automatic recovery
Safety Specifications
Safety Standards
- IT / Multimedia Equipment
- Medical Equipment
- Certification Documents
Pollution Degree
EN 60950-1
EN 62368-1
IEC 60950-1
IEC 62368-1
UL 60950-1
UL 62368-1
EN 60601-1
IEC 60601-1
ANSI/AAMI ES 60601-1
CSA-C22.2, No 60601-1
2 x MOOP (Means Of Operator Protection)
MOPP (Means Of Patient Protection)
www.tracopower.com/overview/tes2m
PD 2
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
www.tracopower.com
September 20, 2023
Page 2 / 4
TES 2M Series, 2 Watt
EMC Specifications
EMI Emissions
EN 60601-1-2 edition 4 (Medical Devices)
EN 55011 class A (with external filter)
EN 55032 class A (with external filter)
EN 55011 class A (with external filter)
EN 55032 class A (with external filter)
External filter proposal: www.tracopower.com/overview/tes2m
EN 60601-1-2 edition 4 (Medical Devices)
- Conducted Emissions
- Radiated Emissions
EMS Immunity
General Specifications
Relative Humidity
Temperature Ranges
Power Derating
- Operating Temperature
- Case Temperature
- Storage Temperature
- High Temperature
Cooling System
Altitude During Operation
Switching Frequency
Insulation System
Working Voltage (rated)
Isolation Test Voltage
Isolation Resistance
Isolation Capacitance
Leakage Current
Reliability
Moisture Sensitivity (MSL)
Washing Process
- Input to Output, 60 s
- Input to Output, 1 s
- Input to Output, 500 VDC
- Input to Output, 100 kHz, 1 V
- Touch Current
- Calculated MTBF
Housing Material
Pin Material
Pin Foundation Plating
Pin Surface Plating
Housing Type
Mounting Type
Connection Type
Footprint Type
Soldering Profile
Weight
Environmental Compliance - REACH Declaration
- RoHS Declaration
- SCIP Reference Number
95% max. (non condensing)
–25°C to +80°C
+105°C max.
–50°C to +125°C
2.22 %/K above 60°C
See application note: www.tracopower.com/overview/tes2m
Natural convection (20 LFM)
4'000 m max.
50 - 100 kHz (PFM)
80 kHz typ. (PFM)
Reinforced Insulation
300 VAC
4'000 VAC
6'000 VAC
10'000 MΩ min.
15 pF typ.
20 pF max.
2 µA max.
2'000'000 h (MIL-HDBK-217F, ground benign)
Level 2 (J-STD-033C)
According to Cleaning Guideline
www.tracopower.com/info/cleaning.pdf
Non-conductive Plastic (UL 94 V-0 rated)
Phosphor Bronze (C5191)
Nickel (5 - 7 µm)
Tin (5 - 7 µm), matte
Plastic Case
PCB Mount
SMD (Surface-Mount Device)
SMD16
Lead-Free Reflow Soldering (acc. J-STD-020E)
See application note: www.tracopower.com/info/reflow-soldering.pdf
3.75 g
www.tracopower.com/info/reach-declaration.pdf
REACH SVHC list compliant
REACH Annex XVII compliant
www.tracopower.com/info/rohs-declaration.pdf
Exemptions: 7a
(RoHS exemptions refer to the component
concentration only, not to the overall
concentration in the product (O5A rule).)
fbb30ad2-edff-45f0-9c69-c10e6eb7222d
Supporting Documents
Overview Link (for additional Documents)
www.tracopower.com/overview/tes2m
All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated.
www.tracopower.com
September 20, 2023
Page 3 / 4
TES 2M Series, 2 Watt
Outline Dimensions
24.0
Pinout
(0.039)
10.6
vent hole with membrane on bottom side
(Ø 0.06)
(to allow proper washing process)
1
7
Dual
1
–Vin (GND)
–Vin (GND)
7
NC
NC
NC
Common
0.15 S
9
+Vout
+Vout
10
–Vout
–Vout
16
+Vin (Vcc)
+Vin (Vcc)
NC: Not connected
(0.71)
top view
Single
8
9
(0.54)
10
13.7
9.4
16
Pin
0-4°
Ø1.5
(0.42)
(0.37)
S
18.1
9.0
1.00
(0.010)
0.25
(0.35)
(0.94)
8
3.11
15.24
2.54
(0.122)
(0.600)
(0.100)
Dimensions in mm (inch)
Tolerances: x.x ±0.5 (x.xx ±0.02)
x.xx ±0.25 (x.xxx ±0.01)
Pin tolerances: ±0.05 (±0.002)
Recommended Solder Pad Layout
1.6
3.11
15.24
2.54
(0.12)
(0.60)
(0.10)
(0.11)
(0.54)
(0.65)
13.7
16.6
2.9
(0.06)
Dimensions in mm (inch)
© Copyright 2023 Traco Electronic AG
Specifications can be changed without notice.
www.tracopower.com
Rev. September 20, 2023
Page 4 / 4
很抱歉,暂时无法提供与“TES 2-2413M”相匹配的价格&库存,您可以联系我们找货
免费人工找货